Clinical and cost-effectiveness of left ventricular assist devices as destination therapy for advanced heart failure: systematic review and economic evaluation
Beese S, Avsar TS, Price M, Quinn D, Lim HS, Dretzke J, Ogwulu CO, Barton P, Jackson L, Moore D
Record ID 32018013224
English
Authors' objectives:
Selected patients with advanced heart failure ineligible for heart transplantation could benefit from left ventricular assist device therapy as ‘destination therapy’. There is evidence of the efficacy of destination therapy; however, it is not currently commissioned within the United Kingdom National Health Service due to the lack of economic evidence. What is the clinical and cost-effectiveness of a left ventricular assist device compared to medical management for patients with advanced heart failure ineligible for heart transplantation (destination therapy)? Heart failure is a debilitating, progressive syndrome characterised by the inability of the heart to pump blood around the body. Pharmacological treatments are used as first-line treatment but may eventually become less effective and left ventricular assist devices (LVADs) or heart transplant (HT) are considered. LVADs are frequently used as bridge to transplant (BTT) or bridge to candidacy (BTC). However, some patients are ineligible for HT and either continue on medical management (MM) or could have a LVAD implanted as ‘destination therapy’ (DT). LVAD as DT is not currently commissioned within the United Kingdom National Health Service (UK NHS). The costs of LVADs are high, especially when compared to the alternative MM, but may also offer significant benefit in terms of survival. It is important to determine whether LVADs are both clinically and cost-effective as DT to inform decision-making from the UK NHS/personal social service (PSS) perspective on their potential as long-term treatment for advanced heart failure patients ineligible for HT.
Authors' results and conclusions:
The clinical effectiveness review included 134 studies (240 articles). There were no studies directly comparing HeartMate 3 and medical management (a randomised trial is ongoing). The currently available left ventricular assist device improves patient survival and reduces stroke rates and complications compared to earlier devices and relative to medical management. For example, survival at 24 months is 77% with the HeartMate 3 device compared to 59% with the HeartMate II (MOMENTUM 3 trial). An indirect comparison demonstrated a reduction in mortality compared to medical management [relative risk of death 0.25 (95% confidence interval 0.13 to 0.47); 24 months; this study]. The cost-effectiveness review included 5 cost analyses and 14 economic evaluations covering different generations of devices and with different perspectives. The reported incremental costs per quality-adjusted life-year gained compared to medical management were lower for later generations of devices [as low as £46,207 (2019 prices; United Kingdom perspective; time horizon at least 5 years)]. The economic evaluation used different approaches to obtain the relative effects of current left ventricular assist devices compared to medical management from the United Kingdom National Health Service/personal social service perspective. All gave similar incremental cost-effectiveness ratios of £53,496–58,244 per quality-adjusted life-year gained – lifetime horizon. Model outputs were sensitive to parameter estimates relating to medical management. The findings did not materially differ on exploratory subgroup analyses based on the severity of heart failure. Using cost-effectiveness criteria applied in the United Kingdom, left ventricular assist devices compared to medical management for patients with advanced heart failure ineligible for heart transplant may not be cost-effective. When available, data from the ongoing evaluation of HeartMate 3 compared to medical management can be used to update cost-effectiveness estimates. An audit of the costs of medical management in the United Kingdom is required to further decrease uncertainty in the economic evaluation. Systematic review of clinical effectiveness There were 240 articles from 134 studies included in the clinical effectiveness review (5 randomised trials, 1 non-randomised trial, 86 observational studies, reports from 5 registries, 5 ongoing studies and 32 systematic reviews). Of the six trials that were included, only one of these assessed the HM3 and this was in comparison to the previous generation HeartMate II device (MOMENTUM RCT). The majority of HM3 data comes from this trial, with minimal additional data contributions from registry reports or observational studies of single cohorts. The MOMENTUM study was considered as having some concerns regarding risk of bias; however, this was primarily due to the per-protocol analysis for the DT participants and most other domains were considered low risk. There were 624 DT patients in the MOMENTUM 3 trial in total, with a mean age of 63 [standard deviation (SD) 12], 82.2% male and 52.1% INTERMACS level 3. At the longest follow-up point (24 months) survival was 76.7% in HM3 DT patients compared to 59% in HeartMate II patients. Clear and significant improvements in QoL from baseline were reported at 12 months and maintained at 24 months in the HM3 group; however, this was similar in the HeartMate II group. Major events and complications were present in both groups by the 24-month follow-up. There were eight stroke events per 100 patient-years in the HM3, as well as one pump thrombosis event and 70 bleeding events per 100 patient-years. These were all lower than that of the HeartMate II group. Rehospitalisations were also significantly lower in HM3 patients. While some reports included HM3 patients, there were no HM3 specific data reported in any patient registry reports. One observational study reported that HM3 patients (n = 15) had 0 pump thrombosis events in 24 months of follow-up. While it was not the focus due to withdrawal, survival levels were lower in the HeartWare HVAD trials when compared to the HM3 in the MOMENTUM trial and there were concerns with the stroke rates reported in the evidence. Risk of bias across the included trials varied with all but two studies reporting an overall high risk of bias or with some concerns for at least one outcome. The evidence contained within the remaining trials, observational studies and registry reports mostly relate to devices other than HM3. This evidence is summarised in the main part of this report. Indirect comparison of HeartMate 3 and medical management As there were no studies directly comparing HM3 and MM, indirect comparisons of the trial data were required utilising MM data from the older REMATCH trial (the first RCT comparing the first-generation HeartMate device to MM). Data were available to link through available studies for the survival outcome only. The network meta-analysis demonstrated a reduction in the risk of mortality, relative risk of death of 0.25 [95% confidence interval (CI) 0.13 to 0.47] 24 months, with the HM3 compared to MM. LVADs have significantly improved over time and the currently available HM3 LVAD is considered clinically effective in patients with end-stage heart failure ineligible for transplant, offering survival of over 75% at 2 years of follow-up with reduced complications and major events in comparison to older devices. However, the device compared to MM may not be considered cost-effective when using methods of defining this for end of life in the UK.
Authors' methods:
A systematic review of evidence on the clinical and cost-effectiveness of left ventricular assist devices as destination therapy was undertaken including, where feasible, a network meta-analysis to provide an indirect estimate of the relative effectiveness of currently available left ventricular assist devices compared to medical management. For the systematic reviews, data sources searched (up to 11 January 2022) were Cochrane CENTRAL, MEDLINE and EMBASE via Ovid for primary studies, and Epistemonikos and Cochrane Database of Systematic Reviews for relevant systematic reviews. Trial registers were also searched, along with data and reports from intervention-specific registries. Economic studies were identified in EconLit, CEA registry and the NHS Economic Evaluation Database (NHS EED). The searches were supplemented by checking reference lists of included studies. An economic model (Markov) was developed to estimate the cost-effectiveness of left ventricular assist devices compared to medical management from the United Kingdom National Health Service/personal social service perspective. Deterministic and probabilistic sensitivity analyses were conducted to explore uncertainties. Where possible, all analyses focused on the only currently available left ventricular assist device (HeartMate 3TM, Abbott, Chicago, IL, USA) in the United Kingdom. There was no direct evidence comparing the clinical effectiveness of HeartMate 3 to medical management. Indirect comparisons made were based on limited data from heterogeneous studies regarding the severity of heart failure (Interagency Registry for Mechanically Assisted Circulatory Support score distribution) and possible for survival only. Furthermore, the cost of medical management of advanced heart failure in the United Kingdom is not clear. Systematic review of clinical effectiveness A systematic review was undertaken of all LVADs as DT and reporting followed the general principles of the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement. The review was registered on PROSPERO (CRD42020158987). Eligibility criteria Studies of patients over 16 years of age with AHF who received any type of LVAD as DT were included. The review considered all devices, but the analysis focused on the HM3 device due to the recent withdrawal of the HeartWare HVAD (Medtronic, Dublin, Republic of Ireland). Eligible comparators (where relevant) were MM and other LVADs. Outcomes were survival, quality of life (QoL), hospitalisations, major events, complications and functional status. Study designs eligible were any clinical trial (whether randomised, non-randomised or single arm), observational studies (cohort, case-controls and case series) and reports from patient registries [e.g. INTERMACS, International Registry for Mechanically Assisted Circulation (IMACS)]. Studies were eligible if 50 or more DT patients were included. Systematic reviews were included and used to identify any additional potentially relevant primary studies.
Authors' identified further research:
Currently, no RCT has been published that compares the HM3 device to MM; however, there is an ongoing trial (SweVAD) comparing the two, which is due to complete final study follow-up in December 2023. This randomised trial, undertaken in Sweden, should allow for relative effects to be determined between the two interventions. This will ultimately enable more robust data to be used to update the current model, rather than relying upon indirect comparisons with wide uncertainty. However, further issues around the true cost of MM are still present due to the lack of recent data on these costs in the UK. An audit of MM costs in DT patients in the UK would address this. Issues also persist in developing reliable subgroup analyses based on severity profiles to aid identification of whether a LVAD is (more) cost-effective for some groups of DT patients. Future trials and other studies should report results by patient severity profiles (e.g. INTERMACS classification), and if registry/observational studies then also by device implanted.
Details
Project Status:
Completed
URL for project:
https://www.journalslibrary.nihr.ac.uk/programmes/hta/NIHR128996
Year Published:
2024
URL for published report:
https://www.journalslibrary.nihr.ac.uk/hta/MLFA4009
URL for additional information:
English
English language abstract:
An English language summary is available
Publication Type:
Full HTA
Country:
England, United Kingdom
DOI:
10.3310/MLFA4009
MeSH Terms
- Heart Failure
- Heart Transplantation
- Costs and Cost Analysis
- Ventricular Dysfunction, Left
- Prostheses and Implants
- Heart-Assist Devices
Contact
Organisation Name:
NIHR Health Technology Assessment programme
Contact Address:
NIHR Journals Library, National Institute for Health and Care Research, Evaluation, Trials and Studies Coordinating Centre, Alpha House, University of Southampton Science Park, Southampton SO16 7NS, UK
Contact Name:
journals.library@nihr.ac.uk
Contact Email:
journals.library@nihr.ac.uk
This is a bibliographic record of a published health technology assessment from a member of INAHTA or other HTA producer. No evaluation of the quality of this assessment has been made for the HTA database.